1. Home
  2. Publications
  3. First evidence of phase-contrast imaging with...

First evidence of phase-contrast imaging with laboratory sources and active pixel sensors

Abstract:
The aim of the present work is to achieve a first step towards combining the advantages of an innovative X-ray imaging technique-phase-contrast imaging (XPCi)-with those of a new class of sensors, i.e. CMOS-based active pixel sensors (APSs). The advantages of XPCi are well known and include increased image quality and detection of details invisible to conventional techniques, with potential application fields encompassing the medical, biological, industrial and security areas. Vanilla, one of the APSs developed by the MI-3 collaboration (see http://mi3.shef.ac.uk), was thoroughly characterised and an appropriate scintillator was selected to provide X-ray sensitivity. During this process, a set of phase-contrast images of different biological samples was acquired by means of the well-established free-space propagation XPCi technique. The obtained results are very encouraging and are in optimum agreement with the predictions of a simulation recently developed by some of the authors thus further supporting its reliability. This paper presents these preliminary results in detail and discusses in brief both the background to this work and its future developments. © 2007 Elsevier B.V. All rights reserved.
Authors:
A Olivo, CD Arvanitis, SE Bohndiek, AT Clark, M Prydderch, R Turchetta, RD Speller
Journal:
Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Citation info:
581(3):776-782
Publication date:
1st Nov 2007
Full text
DOI